HA
H. Akamatsu
21 records found
1
At SRON, we have been developing X-ray TES micro-calorimeters as backup technology for the X-ray Integral Field Unit (X-IFU) of the Athena mission, demonstrating excellent resolving powers both under DC and AC bias. We also developed a frequency-domain multiplexing (FDM) readout
...
We report on the x-ray background rate measured with transition-edge sensors (TES) micro-calorimeters under frequency-domain multiplexing (FDM) readout as a possible technology for future experiments aiming at a direct detection of axion-like particles. Future axion helioscopes w
...
We report our most recent progress and demonstration of a frequency domain multiplexing (FDM) readout technology for transition-edge sensor (TES) arrays, both of which we have been developing in the framework of the X-IFU instrument on board the future Athena X-ray telescope. Usi
...
Large arrays of transition edge sensors (TESs) are the baseline for a number of future space observatories. For instance, the X-ray integral field unit (X-IFU) instrument on board the ATHENA space telescope will consist of ∼ 3000 TESs with high energy resolution (2eV at X-ray ene
...
In the early 2030s, ESAs new X-ray observatory, Athena, is scheduled to be launched. It will carry two main
instruments, one of which is the X-ray Integral Field Unit (X-IFU), an X-ray imaging spectrometer, which will consist of an array of several thousand transition-edge s ...
instruments, one of which is the X-ray Integral Field Unit (X-IFU), an X-ray imaging spectrometer, which will consist of an array of several thousand transition-edge s ...
We present a characterization of the sensitivity of TES X-ray micro-calorimeters to environmental conditions under frequency-domain multiplexing (FDM) readout. In the FDM scheme, each TES in a readout chain is in series with a LC band-pass filter and AC biased with an independent
...
Transition-edge-sensor (TES) microcalorimeters and bolometers are used for a variety of applications.The sensors are based on the steep temperature-dependent resistance of the normal-to-superconducting transition, and are thus intrinsically sensitive to magnetic fields. Conventio
...
We report measured Tc of superconducting Ti/Au bilayer strips with a width W varying from 5 to 50 µm. The strips were fabricated based on a Ti/Au bilayer that consists of a 41-nm-thick Ti layer to which a 280-nm-thick Au layer was added. We find that the Tc
...
Transition-edge sensors (TESs) are the selected technology for future spaceborne x-ray observatories, such as Athena, Lynx, and HUBS. These missions demand thousands of pixels to be operated simultaneously with high energy-resolving power. To reach these demanding requirements, e
...
We report on the development and demonstration of MHz frequency domain multiplexing (FDM) technology to readout arrays of cryogenic transition edge sensor (TES) x-ray microcalorimeters. In our FDM scheme, TESs are AC biased at different resonant frequencies in the low MHz range t
...
Uniform large transition-edge sensor (TES) arrays are fundamental for the next generation of x-ray space observatories. These arrays are required to achieve an energy resolution ΔE < 3 eV full width at half maximum (FWHM) in the soft x-ray energy range. We are currently develo
...
Frequency shift algorithm
Application to a frequency-domain multiplexing readout of x-ray transition-edge sensor microcalorimeters
In the frequency-domain multiplexing (FDM) scheme, transition-edge sensors (TESs) are individually coupled to superconducting LC filters and AC biased at MHz frequencies through a common readout line. To make efficient use of the available readout bandwidth and to minimize the ef
...
Superconducting transition-edge sensors (TESs) are highly sensitive detectors. Based on the outstanding performance on spectral resolution, the X-ray integral field unit (X-IFU) instrument on-board athena will be equipped with a large array of TES-based microcalorimeters. For opt
...
Hot Universe Baryon Surveyor (HUBS), a Chinese space mission, is proposed to find a large fraction of the so-called missing baryons, which would help us to understand more about the structure formation and evolution of the universe. Both theoretical and experimental results show
...
We are developing a transition edge sensor (TES) microcalorimeter array based on a Ti/Au superconducting bilayer, as a backup option for the X-IFU instrument on the Athena X-ray observatory. The array is read out by a frequency-division multiplexing readout system using a 1–5 MHz
...
Dilution and adiabatic demagnetization refrigerators based on pulse tube cryocoolers are nowadays used in many low temperature physics experiments, such as atomic force and scanning tunneling microscopy, quantum computing, radiation detectors, and many others. A pulse tube refrig
...
We are developing X-ray microcalorimeters as a backup option for the baseline detectors in the X-IFU instrument on board the ATHENA space mission led by ESA and to be launched in the early 2030s. 5 × 5 mixed arrays with TiAu transition-edge sensor (TES), which have different high
...
SRON is developing X-ray transition edge sensor (TES) calorimeters arrays, as a backup technology for X-IFU instrument on the ATHENA space observatory. These detectors are based on a superconducting TiAu bilayer TES with critical temperature of 100 mK on a 1 μm thick SiN membrane
...
Frequency-division multiplexing is the baseline read-out system for large arrays of superconducting transition-edge sensors (TES’s) under development for the X-ray and infrared instruments like X-IFU (Athena) and SAFARI, respectively. In this multiplexing scheme, the sensors are
...
SRON is developing ultra-low-noise transition edge sensors (TESs) based on a superconducting Ti/Au bilayer on a suspended SiN island with SiN legs for SAFARI aboard SPICA. We have two major concerns about realizing TESs with an ultra-low NEP of 2×10-19W/Hz: achieving lower therma
...